back
Chamber: Single
More suitable alternative: NO
Produces heat: NO
Produces Static: NO
Low Flash point: NO
Produces Fuel: NO
Produces Adverse: NO
Produces Mechanical: YES
Mechanical Risk - Probability: 1 x Exposure: 1 x Consequence: 1 = Risk Score: 1
The non-flammable liquid is contained in a pump action spray bottle. The technician will require the patient to "burp" the spray bottle during pressurization changes.
See table below to understand risk level corresponding to each risk score.
Risk Score Risk Level Risk Description
> 100 5 Extremely dangerous: do not operate facility
50 - 100 4 Very high: stop use of equipment, processs or procedure
20 - 50 3 High: required urgent attention; proceed with care
5 - 20 2 Medium: attention needed but operation may continue
< 5 1 Low: acceptable risk but noted
* Burman F. Risk Assessment Guide for Installation and Operation of Clinical Hyperbaric Facilities. San Antonio, Texas: International ATMO; 6th Edition, 2019. ISBN: 978-0-578-22098-7

Created 2/12/2025 5:06:23 PM by Marc Wong   MON10412700.pdf