back
Chamber: Single
More suitable alternative: NO
Produces heat: NO
Produces Static: NO
Low Flash point: NO
Iodine has 200 degree flashpoint, polyethylene glycol has 238 flash point
Produces Fuel: YES
Risk of Fire and Explosion - Probability: 1 x Exposure: 3 x Consequence: 1 = Risk Score: 3
Contains iodine and poly
Produces Adverse: YES
Physiological Risk - Probability: 1 x Exposure: 3 x Consequence: 2 = Risk Score: 6
chronic iodine absorption could cause upper respiratory symptoms, effecting pts ability to tolerate pressure change in chamber
Produces Mechanical: NO
See table below to understand risk level corresponding to each risk score.
Risk Score Risk Level Risk Description
> 100 5 Extremely dangerous: do not operate facility
50 - 100 4 Very high: stop use of equipment, processs or procedure
20 - 50 3 High: required urgent attention; proceed with care
5 - 20 2 Medium: attention needed but operation may continue
< 5 1 Low: acceptable risk but noted
* Burman F. Risk Assessment Guide for Installation and Operation of Clinical Hyperbaric Facilities. San Antonio, Texas: International ATMO; 6th Edition, 2019. ISBN: 978-0-578-22098-7

Created 7/18/2024 6:13:37 PM by lauren johnston   msdsum00931 iodoflex dressing.pdf