back
Chamber: Single
More suitable alternative: NO
Produces heat: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 3 = Risk Score: 12
Produces Static: YES
Risk of Fire and Explosion - Probability: 3 x Exposure: 4 x Consequence: 3 = Risk Score: 36
Will have fiberglass cast lay directly on Sechrest matress (sheet lifted) to ground it. The grounding of the chamber is tested daily in our clinic and The grounding of patient before every patient treatment.
Low Flash point: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 3 = Risk Score: 12
Produces Fuel: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 4 = Risk Score: 16
Produces Adverse: NO
Produces Mechanical: NO
See table below to understand risk level corresponding to each risk score.
Risk Score Risk Level Risk Description
> 100 5 Extremely dangerous: do not operate facility
50 - 100 4 Very high: stop use of equipment, processs or procedure
20 - 50 3 High: required urgent attention; proceed with care
5 - 20 2 Medium: attention needed but operation may continue
< 5 1 Low: acceptable risk but noted
* Burman F. Risk Assessment Guide for Installation and Operation of Clinical Hyperbaric Facilities. San Antonio, Texas: International ATMO; 6th Edition, 2019. ISBN: 978-0-578-22098-7

Created 6/6/2024 5:59:33 PM by FMC HBOT   Mediawebserver Scotchcast Plus.pdf