back
Chamber: Single
More suitable alternative: NO
Produces heat: YES
Risk of Fire and Explosion - Probability: 2 x Exposure: 4 x Consequence: 1 = Risk Score: 8
Per manufacturer recommendations re: flammability "Uncured polyurethane tape will burn over an open flame. Finished casts using TECHFORM cast tape will not support combustion under normal conditions as defined by NFPA Standard #702-1975 and the Flammable fabrics Act, CS-191-53"
Produces Static: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 2 = Risk Score: 8
Low Flash point: No Info
Risk of Fire and Explosion - Probability: 2 x Exposure: 4 x Consequence: 1 = Risk Score: 8
Per manufacturer recommendations re: flammability "Uncured polyurethane tape will burn over an open flame. Finished casts using TECHFORM cast tape will not support combustion under normal conditions as defined by NFPA Standard #702-1975 and the Flammable fabrics Act, CS-191-53"
Produces Fuel: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 1 = Risk Score: 4
Produces Adverse: NO
Produces Mechanical: NO
Cast will be covered with Pillow case prior to each HBOT
See table below to understand risk level corresponding to each risk score.
Risk Score Risk Level Risk Description
> 100 5 Extremely dangerous: do not operate facility
50 - 100 4 Very high: stop use of equipment, processs or procedure
20 - 50 3 High: required urgent attention; proceed with care
5 - 20 2 Medium: attention needed but operation may continue
< 5 1 Low: acceptable risk but noted
* Burman F. Risk Assessment Guide for Installation and Operation of Clinical Hyperbaric Facilities. San Antonio, Texas: International ATMO; 6th Edition, 2019. ISBN: 978-0-578-22098-7

Created 6/5/2023 2:04:33 PM by Wendy Rice   Techform premium Cast Tape.pdf