back
Chamber: Single
More suitable alternative: NO
Produces heat: No Info
Risk of Fire and Explosion - Probability: 3 x Exposure: 4 x Consequence: 1 = Risk Score: 12
Cover with moist gauze to keep dressing cool No SDS sent due to not sure of which product was used
Produces Static: No Info
Risk of Fire and Explosion - Probability: 2 x Exposure: 4 x Consequence: 1 = Risk Score: 8
Cover with moist gauze to keep dressing cool
Low Flash point: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 1 = Risk Score: 4
Cover with moist gauze to keep dressing cool
Produces Fuel: No Info
Risk of Fire and Explosion - Probability: 2 x Exposure: 4 x Consequence: 1 = Risk Score: 8
Cover with moist gauze to keep dressing cool
Produces Adverse: NO
Produces Mechanical: NO
See table below to understand risk level corresponding to each risk score.
Risk Score Risk Level Risk Description
> 100 5 Extremely dangerous: do not operate facility
50 - 100 4 Very high: stop use of equipment, processs or procedure
20 - 50 3 High: required urgent attention; proceed with care
5 - 20 2 Medium: attention needed but operation may continue
< 5 1 Low: acceptable risk but noted
* Burman F. Risk Assessment Guide for Installation and Operation of Clinical Hyperbaric Facilities. San Antonio, Texas: International ATMO; 6th Edition, 2019. ISBN: 978-0-578-22098-7

Created 10/27/2022 8:37:06 PM by Jennifer Perry  
t
-->